ENGLISH
99
KEY
SIMBOLS LIT BESIDES
THE NUMBERS
FUNCTION
DESCRIPTION
with RCD
When pressed repeatedly, the
s
key (RCD sub-function) starts and
stops the record function, which
calculates and stores the maxi-
mum, minimum and mean values.
When the RCD key is held down
the instrument emits a short beep
followed by a long one. This con-
firms that the max., min. and
mean values stored previously
are being erased from the
memory and a new series of
recordings is started which will be
used as the basis on which to cal-
culate and store new max., min.
and mean values.
After resetting of the max., min.
and mean values, the RCD func-
tion has a duration of about
30,000 readings, that is 4 hours.
At the end, the RCD function is
automatically interrupted and the
max. min. and mean values calcu-
lated up till then are maintained.
During RCD mode the Auto
Power Off function is disabled and
the battery symbol flashes at a
frequency of 2 Hz.
Attention:
If the battery symbol does not fla-
sh at a frequency of 2 Hz when
the RCD key is pressed, it means
that the RCD function has stored
more than 30,000 acquisitions
and cannot continue any further.
After having taken not of the max.
min. and mean values calculated
mg/l
°