vPad-Mini OPERATORS MANUAL
Figure 16
An image button located at the top left corner of the UI enables the user to select the leakage test type.
The content of the pick list depends on the standard which has been selected via the Settings Menu of
Section 2.2, as follows:
Standard
Available Tests
AAMI-ES1
Enclosure leakage; earth leakage; patient leakage; isolation test;
point-to-point leakage
NFPA 99
Chassis leakage; lead-to-ground leakage; point-to-point leakage
IEC 60601
AU/NZ 3551
Touch current; earth leakage; patient leakage; MAP test;
point-to-point leakage
IEC 62353
Equipment leakage; applied part leakage; point-to-point leakage
IEC 61010
Touch current; point-to-point leakage; accessible voltage test
VDE701-702
Touch current, earth Leakage
When a item on the pick list is chosen, the list is dismissed. Receptacle polarity is not modified when a new
test is selected from the list, however, neutral/line 2 is restored to normal and ground will also be restored
to normal for tests other than earth leakage or equipment leakage. The schematic on the button at top left is
updated to reflect the new test, and the item label is copied to the text field above the control buttons,
prefixed by the name of the standard.
Installation/Chapter 2
#
Page 15
Содержание vPad-Mini
Страница 1: ...Operating Manual MN 125b vPad Mini vPad Safety Analyzer ...
Страница 2: ......
Страница 4: ......
Страница 8: ...vPad Mini OPERATORS MANUAL Table of Contents Page iv ...
Страница 10: ...vPad Mini OPERATORS MANUAL Overview Chapter 1 Page 2 ...
Страница 52: ...vPad Mini OPERATORS MANUAL Operation Chapter 4 Page 44 ...
Страница 54: ...vPad Mini OPERATORS MANUAL ...