![background image](http://html.mh-extra.com/html/datrend-systems/vpad-mini/vpad-mini_operating-manual_514007019.webp)
vPad-Mini OPERATORS MANUAL
Figure 11
2.5.3
When the central button is pressed a pop-up pick list is then shown which allows the user to
change the type of measurement.
As shown in
Figure 11
, this activity provides two measurements: power cord or PE
resistance and point-to-point resistance. The nomenclature of the pick list depends on the
standard which has been selected via the Settings Menu described in Section 2.2 (AAMI and
NFPA standards measure "power cord" resistance; IEC standards measure "protective
earth" resistance).
2.5.4
When a item on the pick list is chosen, the list is dismissed and the schematic on the central
button is updated to reflect the new test, and the label for the selection is copied to the text
field above the schematic.
The toggle button at bottom right controls the test receptacle. Note that it will be OFF when
testing the power cord, and may be OFF or ON when testing point-to-point resistance.
The button icon includes a representation of a LED indicator, which appears bright green when the
receptacle is ON or grey when the receptacle is OFF.
Installation/Chapter 2
#