In the event of a limits failure when testing ALL, the user will be prompted to continue testing while
identifying the function of the Applied Part. The functions will be presented in a list, filtered by AP Type.
The functions list is located in the /Datrend/vPad-Mini/Lists/ folder in the Functions_list file.
The list contains a B, BF or CF prefix to allow the functions to be filtered. If a function can be more than
one Applied Part Type, an entry must exist for each Type.
When a Mains on Applied Part (MAP) test is encountered during a test (IEC 62353 and AU/NZ 3551), the
user will be prompted to connect the leads according to the requirements of the standard.
Note the
Caution
related to the High Voltage present on the AP during the test.