Datex-Ohmeda S/5 Anesthesia Monitor
Document No. 800 1005-1
56
5.1
Gas mother board
Service Data
Oper State
Internal operation state of the module:
0...9
function performed, if staying on, failure is indicated
10...29
Initialization
30...39
Normal operation state
40...49
Zeroing
50...59
Calibrating
ERR status
Indicates measuring unit malfunction:
GAS:
0
no error
1
error in ACX measuring system
2
error in ACX communication
10
error in ASX measuring system
20
error in ASX communication
40
error in PVX measuring system
80
error in PVX communication
Possible failure source: Gas CPU, ACX, ASX or PVX.
ACX:
0
no error
if not 0, replace ACX unit
ASX:
0
no error
if not 0, replace ASX unit
PVX:
0
no error
if not 0, replace PVX unit
Содержание S/5 B-GAS
Страница 2: ......
Страница 6: ...Datex Ohmeda S 5 Anesthesia Monitor Document No 800 1005 1 iv ...
Страница 25: ...S 5 Airway modules Document No 800 1005 1 19 Figure 11 ACX measuring board block diagram ...
Страница 77: ...APPENDIX A Service check form S 5 Airway modules Document No 800 1005 1 71 APPENDIX A ...
Страница 78: ...Datex Ohmeda S 5 Anesthesia Monitor Document No 800 1005 1 72 ...
Страница 82: ...Datex Ohmeda S 5 Anesthesia Monitor A 4 3 Document No 800 1005 1 ...
Страница 83: ......