S/5 Airway modules
Document No. 800 1005-1
47
4
TROUBLESHOOTING
4.1
Troubleshooting chart
Trouble
Possible Cause/Treatment
No response to breathing
Sampling line or water trap blocked or loose, or improperly attached.
Water trap container full.
Interface cable to monitor disconnected.
See chapter “
Gas sampling system troubleshooting
.”
SENSOR INOP. message
The temperature is too high, check fan and filter at the rear panel.
Communication error, check timeout and bad checksum values at the service menu.
Check Airway Module connection cable and supply voltages.
Check ACX measuring board.
xx ZEROING ERROR-message
Gas zeroing failed. Condensation or residual gases are affecting zero measurement. Allow module to run
drawing room air for half an hour and calibrate again.
AIR LEAK-message
Air leak in sampling system. Probably water trap or the sampling line is not attached properly. Gas zero
valve failure. Pump failure or gas outlet blockage. Supply voltage missing
REPLACE TRAP-message
Flow resistance increased due to residue built-up on water trap membrane. Replace the water trap.
REBREATHING-message
CO
2
concentration in inspiratory air is too high. Possibly CO
2
absorber in ventilator is saturated. Change
the absorber.
OCCLUSION-message
Sampling line or water trap is occluded. Water trap container is full. If occlusion persists check internal
tubing for blockages. Check the power supply voltages.
SELECT AGENT-message
No anesthetic agent is selected though delivery is started. Vaporizer valve is broken. Traces of cleaning or
disinfecting agent in the water trap container affecting the readouts. Replace the water trap.
No response to any gas
Sampling line, water trap, or internal tubing blocked or loose, or improperly attached.
Pressure valve malfunction. Pump failure. Supply voltage missing. Serial communication error. Check
those items.
Sudden increase in gas display
Measuring chamber contamination.
±
15 V supply voltages missing.
Water trap malfunction. Check all internal tubing and the interior of the water trap for occlusions or leaks.
Replace water trap.
Check flow rates.
Abnormally high response to all
gases (or abnormally low) or
sudden occlusion warning
Pressure transducer failure. Exchange the ACX Measuring board.
Random output (resembling
noise)
Chopper motor timing pulses out of sync.
Chopper motor not running, motor faulty or connection loose.
Chopper motor driver transistor C-E open circuit or current limiter short circuit.
Exchange the ACX measuring board.
Strong drift in all gases
Leakage in the sampling line or internal tubing (especially in conjunction with too low readings).
Exchange the ACX measuring board.
No gas waveforms
intermittently.
Gas waveforms are not shown during automatic zeroing, occlusion, or air leak. Check the monitor screen
for possible messages and proceed accordingly.
Содержание S/5 B-GAS
Страница 2: ......
Страница 6: ...Datex Ohmeda S 5 Anesthesia Monitor Document No 800 1005 1 iv ...
Страница 25: ...S 5 Airway modules Document No 800 1005 1 19 Figure 11 ACX measuring board block diagram ...
Страница 77: ...APPENDIX A Service check form S 5 Airway modules Document No 800 1005 1 71 APPENDIX A ...
Страница 78: ...Datex Ohmeda S 5 Anesthesia Monitor Document No 800 1005 1 72 ...
Страница 82: ...Datex Ohmeda S 5 Anesthesia Monitor A 4 3 Document No 800 1005 1 ...
Страница 83: ......