Datex-Ohmeda S/5 Anesthesia Monitor
Document No. 800 1005-1
52
4.5
ASX agent identification unit troubleshooting
NOTE: Please read also troubleshooting section in
User’s Reference Manual
.
CAUTION
The agent identification bench ASX-200 can only be repaired and calibrated only at
the factory.
Trouble
Possible Cause/Treatment
AGENT MIXTURE-message when
calibration gas (Freon) is fed.
Repeat calibration.
If the module contains ASX-100, it is not capable to identify calibration gas R23, therefore the
message. However, the ASX-100 will still calibrate with R23.
No response from ASX.
Communication between ASX unit and Central Unit is lost. ASX bench disconnected or faulty.
Check that the motor is running.
No identification
Is the
ASX software
on the software list
on the service
wiev?
Go to ASX service
view and check peak
normal value
Is ASX motor
running?
Check ASX unit
supply voltages
Check Err status in Gas
Mother board service page.
Fault in serial communicat.,
ReplaceASX board or Gas
Mother Board.
NO
YES
YES
NO
Figure 18
ASX troubleshooting flowchart
Содержание S/5 B-GAS
Страница 2: ......
Страница 6: ...Datex Ohmeda S 5 Anesthesia Monitor Document No 800 1005 1 iv ...
Страница 25: ...S 5 Airway modules Document No 800 1005 1 19 Figure 11 ACX measuring board block diagram ...
Страница 77: ...APPENDIX A Service check form S 5 Airway modules Document No 800 1005 1 71 APPENDIX A ...
Страница 78: ...Datex Ohmeda S 5 Anesthesia Monitor Document No 800 1005 1 72 ...
Страница 82: ...Datex Ohmeda S 5 Anesthesia Monitor A 4 3 Document No 800 1005 1 ...
Страница 83: ......