Datex-Ohmeda S/5 Anesthesia Monitor
Document No. 800 1005-1
34
a) shorten the thick O
2
reference channel tube by 4 centimeters (the tube that is connected
to the upper part of the CO
2
absorber)
b) connect the OM ref. filter to the loose end of the tube
c) fasten the filter to the outermost hole in the tubing plate (at the PVX board side) with a
cable tie (order code 64001)
?
•
Replace the D-fend and the sampling line.
NOTE: Use only Datex-Ohmeda sampling lines in order to ensure proper functioning.
•
Connect the module to the Central Unit with a long gas interface cable and switch the
monitor on.
•
Configure the monitor screen so that all the needed parameters are shown, for example as
follows:
Monitor Setup
- Screen 1 Setup - Waveform Fields - Field 1 - Paw
Field 2 - Flow
Field 3 - Off
Field 4 - O2
Field 5 - AA
Field 6 - CO2
Digit Fields - Field 1 - Gases
5.
Check that the module fan is running.
?
6.
Wait until the message ‘Calibrating gas sensor’ disappears from the screen, then enter the
Service menu:
Monitor Setup
- Install/Service
(password 16-4-34)
- Service
(password 26-23-8)
Take down the information regarding Airway module software.
NOTE: The PVX software string does not appear into the ‘Sw version/Unit id’ -list. Check PVX
software from the sticker that is located on the PVX software EPROM (if the PVX board was
installed originally).
?
7.
Enter the ACX service menu:
Parameters - Gas Unit - ACX
Check that the ‘Timeouts’, ‘Bad checksums’ and ‘Bad c-s by mod’ -values are not increasing
faster than by 50 per second. If one of the values is increasing faster, it indicates a failure in
module bus communication.
?
Содержание S/5 B-GAS
Страница 2: ......
Страница 6: ...Datex Ohmeda S 5 Anesthesia Monitor Document No 800 1005 1 iv ...
Страница 25: ...S 5 Airway modules Document No 800 1005 1 19 Figure 11 ACX measuring board block diagram ...
Страница 77: ...APPENDIX A Service check form S 5 Airway modules Document No 800 1005 1 71 APPENDIX A ...
Страница 78: ...Datex Ohmeda S 5 Anesthesia Monitor Document No 800 1005 1 72 ...
Страница 82: ...Datex Ohmeda S 5 Anesthesia Monitor A 4 3 Document No 800 1005 1 ...
Страница 83: ......