Datex-Ohmeda S/5 Anesthesia Monitor
Document No. 800 1005-1
38
Agent identification option
22.
Check that the ‘ACX_ASX Delay’ -value in the ACX service menu is within 400-800.
If the value is not within the range, readjust the flows and repeat the fall time measurement.
?
•
Enter the ASX service menu:
Gas Unit - ASX
NOTE: The ASX service menu values are not updated with the anesthetic identification unit
ASX-100.
23.
Feed calibration gas. When proper absorption spectrum is shown in the menu check that
the Peak normal value is close to 10.50 (±0.20). Check also that the difference between
the “Peak normal” and “Peak mirror” values is not higher than 0.30.
If the values do not meet the range, repeat the gas calibration.
?
•
Set the AA identification to the automatic mode:
Agents - Select Agent - Auto - Previous Menu
24.
Feed calibration gas (order code 755583) and check that the message 'Desflurane'
appears into the digit field for gases.
NOTE: The ASX-100 is not capable of identifying Desflurane.
NOTE: The ASX-100 with software 878364-1.1, or lower, is not capable of identifying
calibration gas R23 (order code 755582). Therefore, the message ‘Agent mixture’ should
appear instead.
?
Содержание S/5 B-GAS
Страница 2: ......
Страница 6: ...Datex Ohmeda S 5 Anesthesia Monitor Document No 800 1005 1 iv ...
Страница 25: ...S 5 Airway modules Document No 800 1005 1 19 Figure 11 ACX measuring board block diagram ...
Страница 77: ...APPENDIX A Service check form S 5 Airway modules Document No 800 1005 1 71 APPENDIX A ...
Страница 78: ...Datex Ohmeda S 5 Anesthesia Monitor Document No 800 1005 1 72 ...
Страница 82: ...Datex Ohmeda S 5 Anesthesia Monitor A 4 3 Document No 800 1005 1 ...
Страница 83: ......