S/5 Airway modules
Document No. 800 1005-1
39
Patient spirometry option
•
Switch the monitor off and install the PVX board, then switch the monitor back on.
•
Preset gas measurement settings:
Airway Gas - Spirometry Setup - Paw Scale - 20
Flow Scale - 15
1.
Check that the patient spirometry connectors on the front panel are clean and intact.
?
2.
Connect a clean spirometry tube to the module and a clean D-lite to the other end of the
tube. Block the D-lite’s sampling line port, for example with a luer stopper.
Take the D-lite into your hand and occlude both ends tightly with your fingers (or with both
hands). Pressing firmly with the fingers creates a pressure inside the D-lite. Check that a
pressure of at least 5 cmH
2
O is generated.
NOTE: If the module has the male & female patient spirometry connectors (pediatric
option), make sure that the date marking on the D-lite is 10/94 or newer.
If the system leaks heavily, no pressure will be generated.
If there is a small leak in the connections, the monitor will measure a pressure difference
which is then interpreted as flow and seen on the monitor screen. The pressure waveform
decreases slowly and the flow waveform either goes above, or below the zero line,
depending on which of the connectors leaks.
In case of leakage, check all connections and try again.
?
3.
Remove the blockage from the sampling line port and connect the sampling line. Breath
through the wider side of the D-lite. Check that the flow waveform moves downwards when
you breath in, and upwards when you breath out.
If the flow waveform moves in opposite manner, check the order of the PVX tubes inside the
module.
?
4.
If possible, check the patient spirometry measurement also with the spirometry tester (order
code 884202). Follow the instructions that are supplied with the tester.
?
Содержание S/5 B-GAS
Страница 2: ......
Страница 6: ...Datex Ohmeda S 5 Anesthesia Monitor Document No 800 1005 1 iv ...
Страница 25: ...S 5 Airway modules Document No 800 1005 1 19 Figure 11 ACX measuring board block diagram ...
Страница 77: ...APPENDIX A Service check form S 5 Airway modules Document No 800 1005 1 71 APPENDIX A ...
Страница 78: ...Datex Ohmeda S 5 Anesthesia Monitor Document No 800 1005 1 72 ...
Страница 82: ...Datex Ohmeda S 5 Anesthesia Monitor A 4 3 Document No 800 1005 1 ...
Страница 83: ......