Datex-Ohmeda S/5 Anesthesia Monitor
Document No. 800 1005-1
50
4.3
OM measuring unit troubleshooting
Because of the complex and very sensitive construction of the oxygen measuring unit no repairs
should be attempted inside the unit. Instead, if the fault has been found in the measuring unit
itself, it should be replaced and the faulty unit be sent to Datex-Ohmeda for repair.
In cases of no response to O
2
or strong drift, check the tubing for loose connections, blockages and
leaks.
CAUTION
Never apply overpressure to the O
2
measuring unit as the pressure transducer may
be permanently damaged.
If the message ‘O
2
zero error’ is displayed check the O
2
measuring unit output voltage on Gas
mother board (see Section Offset adjustment).
If the adjustment range of the (software) calibration is insufficient check the O
2
measuring unit
output voltage and adjust the gain if necessary (see Section Gain adjustment).
If there are problems with O
2
response time check the O
2
measurement flow rate and adjust it if
necessary (see Section Gas Sampling System Adjustments).
If the O
2
signal is noisy, check the measurement unit suspension. Frequency adjustment may help
in some cases (see Section Frequency adjustment).
4.4
ACX troubleshooting
CAUTION
The measuring unit ACX-200 can be repaired and calibrated only at the factory.
The ACX troubleshooting is carried out in the General Troubleshooting scheme. The ACX testing is
explained at the ACX service information section, please refer to it.
4.4.1
Cleaning the measuring chamber of ACX measuring unit
In case the N
2
O “zero calibration constant” in Gas Service (ULT) or N
2
O “calib zero” in ACX Service
(AM/CCM) indicates contamination of the measuring chamber (value 61000 or more), or if the
software will stuck to the “calibrating gas sensor” state due to contamination, it is possible to
attempt the measuring chamber cleaning. However, Datex-Ohmeda recommend exchanging the
measuring unit rather than cleaning.
The measuring chamber should first try to be cleaned by rinsing it with distilled water only. If rinsing
alone is not sufficient the cleaning can be tried with Datex-Ohmeda Cleaning Fluid (85969).
NOTE: Do not use other cleaning agents such as blood gas electrode cleaners. These may damage
the unit.
Содержание S/5 B-GAS
Страница 2: ......
Страница 6: ...Datex Ohmeda S 5 Anesthesia Monitor Document No 800 1005 1 iv ...
Страница 25: ...S 5 Airway modules Document No 800 1005 1 19 Figure 11 ACX measuring board block diagram ...
Страница 77: ...APPENDIX A Service check form S 5 Airway modules Document No 800 1005 1 71 APPENDIX A ...
Страница 78: ...Datex Ohmeda S 5 Anesthesia Monitor Document No 800 1005 1 72 ...
Страница 82: ...Datex Ohmeda S 5 Anesthesia Monitor A 4 3 Document No 800 1005 1 ...
Страница 83: ......