7.10. The number of averaged measurements
(08. menu item)
Function
The instrument performs cca. 12 measurements in each
seconds (5 measurements in 3 wire measuring mode).
The displayed measurement result is generated as the
average of several measurements.
Here you can define the number of measurements that
should be used for calculating the averaged numerical
value. By increasing this number the display stability
increases, but the signal settling time becomes lower.
[Default factory setting: 16]
Sequence of operations
1. Enter the menu with the supervisor code.
The way the code should be type in can be found in
Chapter
7.1. Typing the code in
. You see on the display:
.
2. Keep stepping by pressing the
▲
button as long as you
see this menu item:
(
Ave
raging = átlagolás).
3. Enter the menu item by pressing the
ENTER
button.
4. You see this
on the display. The mnemonic is
blinking it signs that it can be modify.
5. Pressing the
▲
button select the number of
measurements to be averaged for the displayed numerical
value from the followings: 32, 64, 1, 2, 4, 8, 16. [Factory
default: 16].
Note 1: When you choose 1 the displayed value equals to
tle last measured value.
Note 2: the settlig time after stabilizing the input current =
= number of averaging / 12 [seconds].
Number of averaging: 1
Settling time: 0.1 second
Number of averaging: 2
Settling time: 0.2 second
Number of averaging: 4
Settling time: 0.4 second
Number of averaging: 8
Settling time: 0.7 second
Number of averaging: 16
Settling time: 1.4 second
Number of averaging: 32
Settling time: 2.7 second
Number of averaging: 64
Settling time: 5.4 second
The data given above concern for 4 wire or 2 wire mode. In
case of 3 wire mode the settling times must multiply by 2.5.
Содержание DT4120 Series
Страница 1: ...DT4120 xx xx PS Temperature Resistance Potentiometer Meter Transmitter Operating Instructions...
Страница 84: ...DT4120 xx xx PS 84 20170324 V0 10 2 Application example Analog output option...
Страница 85: ...DT4120 xx xx PS 20170324 V0 85 RS232 communication interface option...
Страница 86: ...DT4120 xx xx PS 86 20170324 V0 RS485 communication interface option...
Страница 87: ...DT4120 xx xx PS 20170324 V0 87 RS485 bus topology...
Страница 90: ...DT4120 xx xx PS 90 20170324 V0...
Страница 91: ...DT4120 xx xx PS 20170324 V0 91...
Страница 92: ......