In this example, the Arb type is Voltage.
– Set the Arb Type to Voltage.
– Set V0, V1,t0, t1, t2, t3, t4, Voltage after Arb, and repeat count accordingly. Use the navigation keys to select
the field; use the numeric entry keys to enter the value. The value is set when you press [Enter].
– Set Trigger Source to Key. Press Properties > Trig Src Key.
– Press the color-coded [On] key to enable the output.
– Press Run to start the Arb operation. To abort the operation, press Stop. Alternatively, press [List Run/Stop].
– Press [Scope/Datalog] to enter Scope view and display the Arb.
From the remote interface:
Select the Arb Type and Shape:
ARB:FUNC:TYPE VOLT
ARB:FUNC:SHAP TRAP
Enter the values for the level before and during the top of a voltage trapezoid:
ARB:VOLT:TRAP:STAR 0
ARB:VOLT:TRAP:TOP 5
Enter the time before the trapezoid, the rise-time,fall-time, top-time, and the time after the trapezoid:
ARB:VOLT:TRAP:STAR:TIM 0.25
ARB:VOLT:TRAP:RTIM 0.5
ARB:VOLT:TRAP:FTIM 0.5
ARB:VOLT:TRAP:TOP:TIM 1.5
ARB:VOLT:TRAP:END:TIM 0.25
The parameter setting returns to the DC value that was in effect prior to the Arb.
ARB:TERM:LAST OFF
The parameter setting remains at the last Arb value after the Arb completes.
ARB:TERM:LAST ON
Specify the repeat count of the Arb, or specify continuous repetition (INF):
ARB:COUN 10
ARB:COUN INF
To set the voltage mode to Arb:
VOLT:MODE ARB
Set the trigger source to Key/Immediate:
TRIG:SOUR IMM
To enable the output:
OUTP ON
To initiate the transient trigger system:
INIT:TRAN
98
Keysight E36150 Series User's Guide