SERVICE
16
MAGNET POWER SUPPLY SYSTEM 8500
AD AUTO ADJUSTMENT (Gain & Offset).
From SW version SCS104
The scaling factors for the AD channels can automatically be adjusted with this setting.
Parameters
OFFSET adjustment.
The OFFSET adjustment can automatically be executed if the value lies between 0 and 255 of the
AD channel resolution.
To activate the automatic OFFSET adjustment following steps must be performed:
-
Select SW2 as above. “setting number14"
-
Select the desired channel number “lever 1 to 5" on SW1
-
Set lever 8 on SW1 to OFF (The SW1 LEDs will display the present Offset value)
-
Ensure that the AD input signal is set to zero. Grounded or turned OFF in any way.
-
Press the Setup Button
-
To restore the Offset value to default factory setting (Zero), please set lever 7 to one before
pressing the Setup Button.
DANFYSIK A/S - DK 2630 TAASTRUP - DENMARK.