SERVICE
15
MAGNET POWER SUPPLY SYSTEM 8500
AUXILIARY SETUP 2.
From SW version SCS109
Special options can be initiated with the auxiliary switch set up 2.
(Enhancement of AUXILIARY SETUP 1)
Parameters:
Default setting after “COLD BOOT”
ADC multiplication:
+1
Polarity status signals
Normal.
CAMAC ON/OFF Control
CAMAC.
- From SW version SCC112
Auto slew rate shape
Cosine.
- From SW version SCC113
Lever 1 will negate all current read back readings. This is useful when connected to a regulation loop
that requires a positive signal feed back (negative signal feed back is normal for System 8500).
Lever 2 will inverse the status bits coming from the output polarity switch. If you encounter that the
polarity sign is shown inversed on older systems, then this switch corrects the problem.
Lever 3 determines who controls the ON/OFF/RESET line when the CAMAC bit is active (custom
interface plugged into P3 and pulling pin C32 low).
Lever 4 determines the shape of Auto Slew Rate function.
DANFYSIK A/S - DK 2630 TAASTRUP - DENMARK.