___________________________________________________________
CXR Larus 80-100-400 Issue 2 July 2006 page
3-52
1. Time interval error (TIE) and maximum time interval error (MTIE) over
selected interval. Measurement resolution will be less than or equal to 0.5
nanosecond.
NOTE:
TIE and MTIE are calculated from phase data processed through a 10 Hz low pass filter.
2. Peak-to-peak wander below 10 Hz over selected intervals; also over
5-minute or 24-hour intervals.
3. Peak-to-peak jitter in 10 Hz to 23 kHz band over 1-second intervals.
NOTE:
The clock recovery circuit used for jitter measurement acts as a single pole low pass
filter for jitter frequencies above 23 KHz.
3.3919 Raw Phase Measurement:
Raw unfiltered phase data for any of the five inputs, 32 phase samples per
selectable observation interval, from 1 to 100,000 seconds in decade steps.
3.3920 Reporting of Performance Data:
Data are accessed by local or remote terminals through the 54550 Information
Management Card in the following modes:
a. Synchronization Monitor Card default (scan) mode:
Continuous scan of active E1 or 2048 kHz inputs, accumulating
measurements for 100 seconds for each input. Accumulated values are
stored separately for the 24 most recent one-hour intervals on a rotating
basis (the last 24 hours are stored at any point). Each complete rollover of
24 one-hour intervals is summarized and stored in totals for each of the
past seven days. Cyclic redundancy checking (CRC) events are broken
down into mildly errored seconds (at least one but fewer than 805 CRC
errors) and severely errored seconds (more than 805 CRC errors). All the
stored data can be retrieved on command at any time.
b. Synchronization Monitor Card manual mode:
Remote selection of the parameters to be measured, the inputs to be
scanned, the observation interval (1 to 100,000 seconds in decade steps),
and the number of consecutive intervals (iterations) to be observed on a
given input. Up to 64K iterations can be programmed. If the number of
iterations is set to -1, the data output will continue until stopped by the
operator. Data for the current interval are sent to the terminals at the end
of each iteration.
c. Phase
data:
Selection of one or two of five inputs, the observation interval (1 to 100,000
seconds in decade steps), and the number of consecutive observation
intervals to be measured on a given input. Consecutive intervals can be
selected, with 32 phase samples per interval.
d. RCV
Framers:
Two framers, one of which can be used in the scan mode while the other is
Содержание 54500-16
Страница 7: ...This page is intentionally left blank...
Страница 28: ..._______________________________________________________________ CXR Larus 80 100 400 Issue 2 July 2006 2 3...
Страница 29: ...CXR Larus 80 100 400 2 4 Issue 1 July 2006 ______________________________________________________________...
Страница 30: ..._______________________________________________________________ CXR Larus 80 100 400 Issue 2 July 2006 2 5...
Страница 31: ...CXR Larus 80 100 400 2 6 Issue 1 July 2006 ______________________________________________________________...
Страница 39: ...This page is intentionally left blank...
Страница 205: ...__________________________________________________________ 5 77...
Страница 214: ...AB 2 This page is intentionally left blank...