Impedo
DUO
User Guide
62
5.7.1.3
Communicating directly with a laptop or PC
The port marked ETH. 1 is used to create a local network. The Impedo Duo will issue
DHCP addresses on this port. This port is primarily used to establish communication
to a Laptop/PC that is not connected to an enterprise network. This means the
laptop/PC needs to be setup to receive its IP configuration through DHCP.
Clicking
Discover
on the Address Book page of the Osprey software connects the
instrument and the laptop without requiring any IP information.
The Impedo Duo serves as a DHCP server on ETH 1
Do not connect Ethernet port 1 to a network with an active
DHCP server, as this may cause the network to stop functioning
properly
The IP address of Ethernet port 1 is fixed at 192.168.7.1