Impedo
DUO
User Guide
47
3.6
Power Cycling with the Reset Switch
The Impedo Duo PDU stays powered even if the main processor is powered off. In
this mode the built-in RTC the PDU and the health sensors receive power. After being
fully charged, the Impedo Duo can be safely stored in this state for up to 3 months.
Power can be removed from the entire instrument by the reset switch. The reset
switch is reachable via a small hole close to the GPS port on the right side of the
instrument (see Section 3.1, Figure 19). A paperclip or similar device is required to
toggle the switch. The instrument is powered when the switch is in the depressed
state. Note that removing power from the instrument also removes power from the
RTC.
Upon start-up after a power cycle the RTC is in an unknown state and the instrument
must first be connected to a valid time source (GPS or NTP) to ensure it measures
correctly.
After power cycling the instrument with the reset switch, the
RTC is in an undefined state – the PQ recording system will
require a valid time source after power up before it can function
correctly
Содержание Impedo DUO
Страница 28: ...Impedo DUO User Guide 28 Figure 5 Back of instrument Figure 6 Right side of instrument GPS USB 1...
Страница 85: ...Impedo DUO User Guide 85 Figure 49 Export harmonic selection...
Страница 109: ...Impedo DUO User Guide 109 Appendices 8 Impedo DUO dimensions...
Страница 110: ...Impedo DUO User Guide 110 8 1 Panel mounting plates 8 2 19 installation plate...
Страница 111: ...Impedo DUO User Guide 111 9 Impedo DUO wiring diagram...
Страница 113: ...Impedo DUO User Guide 113 11 Portable ImpedoDUO...
Страница 117: ...Impedo DUO User Guide 117 Screw driver USB with meter key Ethernet cable...