Crest Ultrasonics Corp.
Manual #9907-103-OJSTD
•
Page 13
2.2 INSTALLATION
The console should be carefully uncrated and inspected
for damage. Follow your facility’s standard procedures for
making notification if any shipping damage is found.
The console should be situated in a location that allows
adequate flow of cooling air
.
The console, because of its size and weight, or if the
plumbing connections are fixed cannot be readily moved.
Crest recommends that the console should be situated at
least 2 feet from the walls to allow space for service
accessibility as well as personnel safety.
Follow assembly instructions in “Reference” if your system
includes a Crest CTS-model materials handling system.
WARNING: Before working with any part of an
ultrasonic cleaning system, you must tightly secure
loose hair, jewelry, loose clothing and other personal
effects that may be drawn into or damaged by moving
parts of the machinery. Failure to do this may result in
property damage and/or serious personal injury.
PLUMBING INSTALLATION
:
The console must be connected to water and air inlet lines,
exhaust duct (for exhaust option), and unrestricted waste
outlet lines. Connections should be done according to
instructions on the following pages and information
provided on the plumbing schematics for this system.
WARNING: Electrical power should be off when
plumbing connections are being done.
Plumbing must be done in accordance with all local and
regional regulations. All environmental safety and
protection guidelines must be observed.
The use of unions in the connections is advisable for
serviceability. It is recommended that you install external
valves for all feed lines connected to the unit.
Movement during shipping may loosen certain plumbing
connections which have already been made by Crest
manufacturing. Therefore, you should check all the
compression fittings and unions on your Crest Ultrasonics