Installation - Prepare the Quark Unit and Subject for Exercise Testing - 37
Turbine Attached
Turbine Not Attached
No Inspiratory Valves
“Inspire”
Inspiratory Valves
“Expire”
Instructions
Wind shield not attached
Hand over open end, occlude and
have subject breathe as shown
Hand over open end, occlude and
have subject breathe as shown
The HR Monitor
Assemble the HR Monitor Belt:
1.
Snap the transmitter onto the elastic strap
2.
Wet the electrode surfaces on the inside of the belt on either side of the transmitter with water, contact lens solution, electrode gel to improve
conductivity
3.
Adjust the strap to fit tightly and comfortably around the subject’s thorax without restricting breathing
4.
Secure the strap tightly around the chest (below the nipple line) and lock the buckle
5.
The HR Monitor must be worn against bare skin to ensure successful operation. Best Practice – To acquire the most accurate HR signal, the HR
transmitter and receiver should be in close proximity