Getting Started - Environmental Conditions for Operation - 13
Environmental Conditions for Operation
1.
COSMED devices should not be operated near explosive substances
2.
COSMED devices should not be installed near electrical or magnetic devices such as x-ray equipment, transformers or power lines. These devices
could create electrical interference when performing testing procedures.
3.
COSMED devices are not AP or APG units (according to EN 60601-1) and should never be operated in the presence of flammable anaesthetic
mixtures
4.
COSMED devices should be operated under normal environmental temperatures and conditions [IEC 60601-1/EN 60601-1]:
•
Temperature
: 10°C (50°F) and 40°C (104°F)
•
Relative humidity
: 30% to 90% (not condensing)
•
Atmospheric Pressure
: 600 mBar (450 mmHg) to 1060 mBar (795 mmHg)
•
Avoid operating equipment in the presence of noxious fumes or in a dusty environment
•
Do not place units near heating sources
•
Cardiopulmonary resuscitation equipment should be accessible in the case of an emergency
•
Adequate floor space and easy access to the subject during exercise testing is necessary
•
Adequate ventilation should be maintained in the room the testing is to be performed
•
No other testing should be performed in the room simultaneously
Содержание Quark CPET
Страница 2: ......
Страница 8: ......
Страница 9: ...Getting Started...
Страница 25: ...Installation...
Страница 43: ...System Cleaning Maintenance...
Страница 48: ...46 System Cleaning Maintenance Spirometry Tests Spirometry Tests...
Страница 56: ...54 System Cleaning Maintenance CPET RMR Tests CPET RMR Tests...
Страница 63: ...System Cleaning Maintenance Canopy test 61 Canopy test...
Страница 68: ...66 System Cleaning Maintenance Mixing Chamber Test Mixing Chamber Test...
Страница 77: ...System Cleaning Maintenance Rocc test 75 Rocc test...
Страница 83: ...Appendix...
Страница 101: ......