92 - Appendix - Technical Specifications
Technical Specifications
Flowmeter
ID28 turbine
ID18 turbine
X9
Type: Bidirectional
Bidirectional
PNT
Lilly
Diameter (int.):
28mm
18mm
28mm
Flow range:
0-16 l/s
0-8 l/s
0-14 l/s
Volume range:
0-300 l/min
0-50 l/min
Resolution:
12 ml
3 ml
1 ml
Accuracy:
± 2% or 20 ml/s
± 2% or 20 ml/s
± 2% or 20 ml/s
Resistance: <0.6
cmH
2
O/l/s <0.7
cmH
2
O/l/s <1
cmH
2
O/l/s
@ 14 l/s
@ 3 l/s
@ 14 l/s
O
2
analyzer
Type: Paramagnetic
Response time:
120 ms
Range:
0-30% (0-100% FRC)
Accuracy: ±0.1%
Resolution:
0.01% (0.03% FRC)
Warm-up time:
5 min
CO
2
analyzer
Type: Digital
infrared
Response time:
100 ms
Range: 0-10%
Accuracy: ±0.1%
Resolution: 0.01%
Warm-up time:
10 min
CO analyzer
Type: Infrared
Response time:
200 ms
Range: 0-0.35%
Accuracy: ±0.003%
Resolution: 0.001%
Warm-up time:
15 min
CH
4
analyser
Type: Infrared
Response time:
200 ms
Range: 0-0.35%
Accuracy: ±0.003%
Resolution: 0.001%
Warm-up time:
15 min
Humidity absorber
Capillary of Nafion (Permapure ®)
Power Supply
Voltage:
100V-240V ±10%; 50/60Hz
Power consumption:
100VA
Environmental Sensors
Temperature: 0-50°C
Barometer: 400-800
mmHg
Humidity: 0-100%
Dimension and Weight
Dimensions: 16x33x41
cm
Weight:
11 Kg (weight depends on the
configuration)
Oximeter
A
RMS
(70%-100%)
1.70
A
RMS
(60%-70%)
2.16
A
RMS
(70%-80%)
1.90
A
RMS
(80%-90%)
1.80
A
RMS
(90%-100%)
1.34
Accuracy specifications
FVC
±3.5% or 0.100L/s whichever is greater
FEV1
±3.5% or 0.100L/s whichever is greater
FEF25-75%
±5.5% or 0.250L/s whichever is greater
PEF
±7% or ±0.420 L/s whichever is greater
MVV
±10.5% or 20 L/min whichever is greater
FRC ±5%
DLCO ±5%
MIP/MEP ±3%
P0.1
±5% or 0.5 cmH
2
O whichever is greater
SpO
2
±1%
Ve ±3%
RF ±3%
HR ±2
units
VO
2
±3%
VCO
2
±3%
Содержание Quark CPET
Страница 2: ......
Страница 8: ......
Страница 9: ...Getting Started...
Страница 25: ...Installation...
Страница 43: ...System Cleaning Maintenance...
Страница 48: ...46 System Cleaning Maintenance Spirometry Tests Spirometry Tests...
Страница 56: ...54 System Cleaning Maintenance CPET RMR Tests CPET RMR Tests...
Страница 63: ...System Cleaning Maintenance Canopy test 61 Canopy test...
Страница 68: ...66 System Cleaning Maintenance Mixing Chamber Test Mixing Chamber Test...
Страница 77: ...System Cleaning Maintenance Rocc test 75 Rocc test...
Страница 83: ...Appendix...
Страница 101: ......