48 - System Cleaning & Maintenance - Spirometry Tests
Item Turbine
PNT
X9
When?
Before first use and immediately after use
Daily
Notes
1.
Do not place the turbine under running water or
move the turbine while submerged
2.
Do not use a brush for the internal part of the turbine,
doing so will damages to the turbine blade
3.
For cleaning and rinsing, do not wet the sampling
line.
4.
After cleaning the turbine, always calibrate prior to
subsequent testing.
1.
Use distilled water for preparing the disinfectant
solution, otherwise calcium deposits could damage
the flowmeter mesh
2.
Do not dry the PNT X9 with hot air which may damage
the flowmeter net (<40°C)
3.
This item is protected by the antibacterial filter
Pretreatment
Rinse the component in tap water at 22-40°C
Rinse the component in tap water at 22-40°C
Manual
Cleaning/Rinsing
1.
Soak for 3 minutes in a mild detergent solution
(Cidezyme) at room temperature (>22°C). Using a
soft bristle brush, scrub the item submerged in the
cleaning solution during the 3 minutes (only scrub
outside only)
2.
Wipe while rinsing for 5 minutes in a large volume of
water. Using a lint free towel or gauze, remove the
water in excess from the item
3.
Inspect all surfaces and item features to ensure that it
is visibly clean, repeat Manual Cleaning if not
1.
Soak for 3 minutes in a mild detergent solution
(Cidezyme) at room temperature (>22°C). Using a
soft bristle brush, scrub the item submerged in the
cleaning solution during the 3 minutes
2.
Wipe while rinsing for 5 minutes in a large volume of
water. Using a lint free towel or gauze, remove the
water in excess from the item
3.
Inspect all surfaces and item features to ensure that it
is visibly clean, repeat Manual Cleaning if not
Disinfection
1.
Soak the component in disinfectant (Cidex OPA) at
room temperature (>22°C) for 12 minutes
2.
Rinse three (3) times in a large volume of water for at
least 3-minutes for each rinsing cycle
3.
Air dry
1.
Soak the component in disinfectant (Cidex OPA) at
room temperature (>22°C) for 12 minutes
2.
Rinse three (3) times in a large volume of water for at
least 3-minutes for each rinsing cycle
3.
Air dry
Cleaning Time
30 minutes
30 minutes
Drying Time
2-3 hours
2-3 hours
Содержание Quark CPET
Страница 2: ......
Страница 8: ......
Страница 9: ...Getting Started...
Страница 25: ...Installation...
Страница 43: ...System Cleaning Maintenance...
Страница 48: ...46 System Cleaning Maintenance Spirometry Tests Spirometry Tests...
Страница 56: ...54 System Cleaning Maintenance CPET RMR Tests CPET RMR Tests...
Страница 63: ...System Cleaning Maintenance Canopy test 61 Canopy test...
Страница 68: ...66 System Cleaning Maintenance Mixing Chamber Test Mixing Chamber Test...
Страница 77: ...System Cleaning Maintenance Rocc test 75 Rocc test...
Страница 83: ...Appendix...
Страница 101: ......