58 - System Cleaning & Maintenance - CPET & RMR Tests
Item Flowmeter
Sampling line tip
When?
Before first use and immediately after use
Before first use and immediately after use
Note
Do not clean and disinfect the internal surface of the
reader
Lubricate the O-ring of the sampling line plug after
cleaning with biocompatible lubricant
Pretreatment
Wipe soil with a moist sponge or lint-free towel
Wipe soil with a moist sponge or lint-free towel
Manual Cleaning
Wipe with a damp soft cloth
Wipe with a damp soft cloth
Disinfection
Wear recommended PPE according to the
bottle. Remove one Clinell Sporicidal
wipe from the pack
To activate, wet the wipe with water
under a tap or soak in a bucket. Squeeze
out the wipe to remove excess water
Wipe the surface, taking care not to go
over the same area twice
Change wipe if it becomes dry or soiled
and discard in the clinical waste bin. Let
the surface air dry
Wear recommended PPE according to the
bottle. Remove one Clinell Sporicidal
wipe from the pack
To activate, wet the wipe with water
under a tap or soak in a bucket. Squeeze
out the wipe to remove excess water
Wipe the surface, taking care not to go
over the same area twice
Change wipe if it becomes dry or soiled
and discard in the clinical waste bin. Let
the surface air dry
Cleaning Time
1 minutes
1 minutes
Drying Time
10-15 minutes
10-15 minutes
Содержание Quark CPET
Страница 2: ......
Страница 8: ......
Страница 9: ...Getting Started...
Страница 25: ...Installation...
Страница 43: ...System Cleaning Maintenance...
Страница 48: ...46 System Cleaning Maintenance Spirometry Tests Spirometry Tests...
Страница 56: ...54 System Cleaning Maintenance CPET RMR Tests CPET RMR Tests...
Страница 63: ...System Cleaning Maintenance Canopy test 61 Canopy test...
Страница 68: ...66 System Cleaning Maintenance Mixing Chamber Test Mixing Chamber Test...
Страница 77: ...System Cleaning Maintenance Rocc test 75 Rocc test...
Страница 83: ...Appendix...
Страница 101: ......