File : CS-SUP-MUTI-WTHERMAPP-E01.sxw
Contents of request REQ_SEND_FRAME
Data Field (max : 152 bytes)
Applicative Command
Precision level
1 byte
1 byte
0x01
(*)
(*) The precision field is used only for the modules :
- WaveTherm – PT100
- WaveTherm – PT1000
The WaveTherm – DALLAS doesn't support the precision capability. In this case the precision field
doesn't exist (size = 0 byte).
Thus, the format of the reponse varies according to the module used :
➢
WaveTherm – DALLAS :
Contents of the response RECEIVED_FRAME
Data Field (max : 152 bytes)
Acknowledgement
of the Applicative
Command
Operating
Mode
Application
Status
MSB
Temperature
A
LSB
Temperature
A
MSB
Temperature
B
LSB
Temperature
B
1 byte
1 byte
1 byte
1 byte
1 byte
1 byte
1 byte
0x81
(*)
(*)
(*)
(*)
(*)
These fields are set only if the sensors are wired, and initialized, otherwise the returned value is 0x4FFF.
The format of the temperature is described in chapters 5.1.1 and 5.2.
➢
WaveTherm – PT100 and WaveTherm – PT1000 :
Contents of the response RECEIVED_FRAME
Data Field (max : 152 bytes)
Acknowledgement
of the Applicative
Command
Operating
Mode
Application
Status
Temperature A
(Float 32 bits
LSB First)
Temperature B
(Float 32 bits
LSB First)
1 byte
1 byte
1 byte
4 bytes
4 bytes
0x81
(*)
(*)
These fields are set only if the sensors are wired, otherwise the returned value is 0xFFFFFFFF.
WaveTherm modules – application handbook