S2VNA and S4VNA SCPI Programming Manual
257
SENS:CORR:COLL:LOAD
SENSe
<Ch>
:CORRection:COLLect[:ACQuire]:LOAD
<port>
Description
Measures the calibration data of the
load
standard for the specified
port.
(no query)
Target
Channel
<Ch>
,
<Ch>
={[1]|2|...16}
Parameter
<
port
>
the number of the port from 1 to 2 (or 4)
Notes
The command starts the measurement immediately if the trigger source
for
calibration
set
to
the
"Internal"
by
the
command
SENS:CORR:TRIG:FREE, otherwise waits for the trigger signal. The
command blocks the execution of the subsequent commands until the
completion of the measurement.
Equivalent Softkeys
Calibration > Calibrate > Response (Open) > Load (Optional)
Calibration > Calibrate > Response (Short) > Load (Optional)
Calibration > Calibrate > Full 1
–
Port Cal > Load
Calibration > Calibrate > One Path 2
–
Port Cal > Load
Calibration > Calibrate > Full 2
–
Port Cal > Port n Load