S2VNA and S4VNA SCPI Programming Manual
261
SENS:CORR:COLL:TRLL
SENSe
<Ch>
:CORRection:COLLect[:ACQuire]:TRLLine
<port1>
,
<port2>
Description
Measures the calibration data of the TRL line
standard between the
<port1>
and the
<port2>
.
(no query)
Target
Channel
<Ch>
,
<Ch>
={[1]|2|...16}
Parameter
<
port1
>
the number of the port from 1 to 2 (or 4)
<
port2
>
the number of the port from 1 to 2 (or 4)
Out of Range
If the same port numbers are specified, an error occurs.
Notes
The command starts the measurement immediately if the trigger source
for
calibration
set
to
the
"Internal"
by
the
command
SENS:CORR:TRIG:FREE, otherwise waits for the trigger signal. The
command blocks the execution of the subsequent commands until the
completion of the measurement.
Equivalent Softkeys
Calibration > Calibrate > n
–
Port TRL Cal > Line/Match