S2VNA and S4VNA SCPI Programming Manual
293
SENS:CORR:EXT:PORT:TIME
SENSe
<Ch>
:CORRection:EXTension:PORT
<Pt>
:TIME
<time>
SENSe
<Ch>
:CORRection:EXTension:PORT
<Pt>
:TIME?
Description
Sets or reads out the electrical delay value for the port extension
function.
(command/query)
Target
Port
<Pt>
of channel
<Ch>
,
<Ch>
={[1]|2|...16}
<Pt>
={[1]|2} for S2VNA or {[1]|2|3|4} for S4VNA
Parameter
<
time
>
the electrical delay value from –10 to 10
Unit
s (second)
Out of Range
Sets the value of the limit, which is closer to the specified value.
Query Response
<numeric>
Preset Value
0
Equivalent Softkeys
Calibration > Port Extensions > { Extension Port1 | Extension Port2 }