S2VNA and S4VNA SCPI Programming Manual
240
SENS:CORR:COLL:CKIT:STAN:L3
SENSe:CORRection:COLLect:CKIT:STAN
<Std>
:L3
<numeric>
SENSe:CORRection:COLLect:CKIT:STAN
<Std>
:L3?
Description
Sets or reads out the L3 value for the short calibration standard.
(command/query)
Target
Standard
<Std>
of the selected calibration kit,
<Std>
={[1]|2|...N}
,
where N – the number of the standards in
the calibration kit
Parameter
<
numeric
>
the L3 value from –1E18 to 1E18
Unit
1E–42 H/Hz
3
(Henry/Hertz
3
)
Out of Range
Sets the value of the limit, which is closer to the specified value.
Query Response
<numeric>
Equivalent Softkeys
Calibration > Cal Kit > Define STDs > L3 10
–
42
H/Hz
3
SENS:CORR:COLL:CKIT:STAN:LAB
SENSe:CORRection:COLLect:CKIT:STAN
<Std>
:LABel
<string>
SENSe:CORRection:COLLect:CKIT:STAN
<Std>
:LABel?
Description
Sets or reads out the label for the calibration standard.
(command/query)
Target
Standard
<Std>
of the selected calibration kit,
<Std>
={[1]|2|...N}
,
where N – the number of the standards in
the calibration kit
Parameter
<
string
>
, up to 254 characters
Query Response
<string>
Equivalent Softkeys
Calibration > Cal Kit > Define STDs > Label
Содержание C1209
Страница 1: ...S2 and S4 series Network Analyzer Programming Manual SCPI Commands Software version 19 1 2019...
Страница 12: ...S2VNA and S4VNA SCPI Programming Manual 12 Appendix 3 Programming Examples 423...
Страница 31: ...S2VNA and S4VNA SCPI Programming Manual 31...
Страница 427: ...427...
Страница 428: ...428...