S2VNA and S4VNA SCPI Programming Manual
342
SENS:VOLT:DC:RANG:UPP
SENSe
<Ch>
:VOLTage{[1]|2}:DC:RANGe:UPPer
<numeric>
SENSe
<Ch>
:VOLTage{[1]|2}:DC:RANGe:UPPer?
Description
Sets or reads out the DC voltage range at the connector AUX1 or
AUX2.
(command/query)
Target
Channel
<Ch>
,
<Ch>
={[1]|2|...16}
Parameter
<
numeric
>
the
DC voltage range 10V or 1V
Unit
V (Volt)
Out of Range
Sets the value of the limit, which is closer to the specified value.
Query Response
<numeric>
Preset Value
10
Equivalent Softkeys
Measurement > DC Voltage > Range
SERV:CHAN:ACT?
SERVice:CHANnel:ACTive?
Description
Reads out the active channel number.
(query only)
Query Response
<
numeric
>
from 1 to 16
Equivalent Softkeys
None