S2VNA and S4VNA SCPI Programming Manual
252
SENS:CORR:COLL:DATA:OPEN
SENSe
<Ch>
:CORRection:COLLect:DATA:OPEN
<port>
,
<numeric list>
SENSe
<Ch>
:CORRection:COLLect:DATA:OPEN?
<port>
Description
Writes or reads out the array of the
open
calibration standard
measurement for the port
<port>
.
The array size is 2N, where N is the number of measurement
points.
For the n–th point, where n from 1 to N:
<numeric 2n
–
1>
real part of the measurement
<numeric 2n>
imaginary part of the measurement
(command/query)
Target
Channel
<Ch>
,
<Ch>
={[1]|2|...16}
Parameter
<
port
>
the number of the port from 1 to 2 (or 4)
<
numeric
list
>
the data array of the
open
standard measurement
Query Response
<numeric 1>
,
<numeric 2>
, …
<numeric 2N>
Related Commands
SENS:CORR:COLL:OPEN
Equivalent Softkeys
None