S2VNA and S4VNA SCPI Programming Manual
305
SENS:CORR:STAT
SENSe
<Ch>
:CORRection:STATe {OFF|ON|0|1}
SENSe
<Ch>
:CORRection:STATe?
Description
Turns ON/OFF the S-parameter error correction.
(command/query)
Target
Channel
<Ch>
,
<Ch>
={[1]|2|...16}
Parameter
{ON|1}
: ON
{OFF|0}
: OFF
Query Response
{0|1}
Preset Value
0
Equivalent Softkeys
Calibration > Correction
SENS:CORR:TRAN:TIME:FREQ
SENSe
<Ch>
:CORRection:TRANsform:TIME:FREQuency
<frequency>
SENSe
<Ch>
:CORRection:TRANsform:TIME:FREQuency?
Description
Sets or reads out the
frequency value at which the cable loss specified
for the cable correction function, when the time domain transformation
function is turned ON
.
(command/query)
Target
Channel
<Ch>
,
<Ch>
={[1]|2|...16}
Parameter
<
frequency
>
the frequency value.
Unit
Hz (Hertz)
Query Response
<numeric>
Preset Value
1 GHz
Equivalent Softkeys
Analysis > Time Domain > Cable Correction > Frequency