59
Firmware upgrade
To upgrade the firmware file, the configuration file and the parameters file, follow these
steps:
1. Connect the steriliser to the computer using the USB serial cable.
2. Turn on the sterilizer from the
main switch
13
.
3. Turn on the computer and install the program SterilClaveVxx.exe following the
instructions displayed on the screen.
4. Start the program SterilClaveVxx.exe.
5. Choose the data transfer rate: select 115200 bps.
6. Type the number of the COM port to which the steriliser is connected.
7. Click Open.
8. Click the Debug tab.
9. Click Identify. If the box below displays numbers different from zero the connection
has been established correctly; otherwise, check the number of the COM port and
the cable because there is something wrong with the connection.
10. Click the Download tab to download the firmware file, the configuration file and
the parameter file, IN THIS ORDER.
11. Click Download Firmware File and select the firmware file (.s19) from a folder to start
downloading. When the download is completed, the message OPERATION
SUCCESSFUL will be displayed in the box.
12. Click Download Configuration File, select the configuration file (.cmxcnf) from a
folder to start downloading.
13. Click Download Parameters File, select the parameters file (.cmxpars) from a folder
to start downloading.
Click Close to close the connection and click Exit to exit the program.
Firmware upgrade
Содержание STERILCLAVE 18 B
Страница 2: ...2...
Страница 6: ...6 Graphical reference representation Model 6 S Dynamica Contents...
Страница 7: ...7 Graphical reference representation Model 18 Contents...
Страница 8: ...8 Graphical reference representation Model 24 Contents...
Страница 9: ...9 Graphical reference representation of common components 40 Contents...
Страница 10: ...10...
Страница 63: ...63 Diagrams and wirings...
Страница 75: ...75...