44
Door adjustment
•
Procedure for 18/24 litre models: if leakages still occur after checking and cleaning
chamber seal
31
and the flange of
chamber
09
on which it rests, you will need to:
-
Switch the steriliser off and wait for it to cool down, then open the door.
-
Use a wrench
ch5
to loosen the screw inside the hole on the left side of the
door. When the screw is loose, the plate will be free to rotate.
-
Turn the plate anticlockwise for about ¼ of a turn or as required.
-
Tighten the Allen screw
ch5
.
-
When closing the door, a higher resistance should be felt.
-
Test tightness by starting a new cycle. Repeat the operation if required.
•
For 6 S Dynamica, in case of leaks do the following:
-
Check and clean
chamber seal
31
.
-
Adjust the door closing by following these steps:
o
Switch the steriliser off and wait for it to cool down, then open the
door.
o
Remove the cover from the steriliser.
o
Rotate the screw on top of the chamber to adjust the hook and thus
adjusting the door closing.
o
Test tightness by starting a new cycle. Repeat the operation if
required.
Note:
Do not tighten the door excessively. Excessive pressure may deform the seal.
Additional Checks And Settings
Содержание STERILCLAVE 18 B
Страница 2: ...2...
Страница 6: ...6 Graphical reference representation Model 6 S Dynamica Contents...
Страница 7: ...7 Graphical reference representation Model 18 Contents...
Страница 8: ...8 Graphical reference representation Model 24 Contents...
Страница 9: ...9 Graphical reference representation of common components 40 Contents...
Страница 10: ...10...
Страница 63: ...63 Diagrams and wirings...
Страница 75: ...75...