ALARMS
Alarms indicate the presence of a possible fault and the need for repairs.
If a fault occurs during a cycle, the cycle will be immediately interrupted and a buzzer will
sound for 15 seconds to indicate the alarm conditions. A description of the alarm will
appear on the display along with the cycle and phase when the alarm occurred. To reset
an alarm, press SILENCE ALARM and enter the security code or
SterilCard
06
.
An alarm will cause the steriliser to be switched off. Therefore, after cancelling the alarm
and solving the cause, you will need to switch the steriliser back on before starting a new
cycle.
The load in the steriliser during the cycle interrupted by the alarm is to be considered NON
STERILE and must be dried and repacked, if required.
Description /
Cause
/
•
Solution
Insufficient water quality alarm
This alarm is available only on models fitting an external water demineralisation system
(optional) or if the water is automatically loaded from an external tank (optional; standard
on BHD models); in this case, the steriliser monitors conductivity of loaded water.
In such conditions, the alarm indicates that the water conductivity value has exceeded an
acceptable level and stops the automatic water inlet.
The steriliser will display this message when a cycle is started (see USE AND MAINTENANCE
MANUAL, chapter ALARMS), only when an automatic water inlet
29
is installed (see USE
AND MAINTENANCE MANUAL, chapter CONNECTIONS) and Self-filling in the Main Settings
Menu is ON.
It indicates that the conductivity of the water provided is no longer acceptable (>30
µS/cm); this alarm is always preceded by the CHECK WATER QUALITY message.
•
If the steriliser is connected to a demineralisation system, change the cartridge
following the manufacturer’s instructions. After replacing the cartridges, rinse the
clean
water tank
10
.
•
If the steriliser takes water from an external tank, change the water because it is no
longer suitable.
•
If you want to run the steriliser without replacing the cartridges in the water
demineralisation unit, re-start the steriliser by setting Self-filling OFF and manually fill the
tank through the
filler
14
. In this case, the water will not be monitored and you should
make sure that the quality is suitable.
Insufficient water level alarm
It indicates that there is no distilled or demineralised water in the
clean water tank
10
.
The steriliser will show this alarm if probe S2 is uncovered when cycle is started or if probe S1
is uncovered during a cycle.
•
If the alarm appears when a cycle is started
,
fill the
clean water tank
10
.
To fill the tank without the self-filling function, loosen and remove the plug from
filler
14
and insert the supplied
measuring funnel
05
(use suitable water only).
If the steriliser is on and in STAND BY when filled, the
“Minimum water level” and
“Maximum water level”
messages on the display may be used to check the level. A
Alarms
Содержание STERILCLAVE 18 B
Страница 2: ...2...
Страница 6: ...6 Graphical reference representation Model 6 S Dynamica Contents...
Страница 7: ...7 Graphical reference representation Model 18 Contents...
Страница 8: ...8 Graphical reference representation Model 24 Contents...
Страница 9: ...9 Graphical reference representation of common components 40 Contents...
Страница 10: ...10...
Страница 63: ...63 Diagrams and wirings...
Страница 75: ...75...