38
•
In configuration (B)
, check that the water supply tap is open and that the mains water
pressure is higher than the minimum pressure indicated for the installed demineralisation
system.
•
If the problem persists, check if solenoid valve
EV7
opens correctly.
In both cases
, check:
- that all connections are properly connected;
- that all pipes are in their seats and neither cracked nor twisted.
Door alarm
This indicates that
door
16
is not correctly closed.
The steriliser may display this message when an attempt is made to start a cycle but the
door is not completely locked. The steriliser will not start in this condition.
•
Close the door well before starting a cycle.
•
If the problem persists, despite the door being locked,
micro-switch M1
36A
and/or
micro-switch M2
36B
are not correctly pressed or faulty. Check that:
- the lock nuts on the micro-switches are tight;
- the pin is not bent, out of its seat, or loose;
- there is continuity at micro-switch contacts (use a tester);
- there are no obstructions or dirt near the micro-switch rod;
- the connection cables between the door and the steriliser are operating and
properly connected.
Note:
To determine which of the two micro-switches is generating the signal, check:
- if the alarm is triggered immediately
after starting the cycle, it means that the pre-closing
micro-switch is not working (
micro-switch M2
36B
)
;
- if the alarm is triggered 3 minutes after
starting the cycle, it means that the door lock
micro-switch is not working (
micro-switch M1
36A
).
Micro-switch M1
36A:
Alarms
Содержание STERILCLAVE 18 B
Страница 2: ...2...
Страница 6: ...6 Graphical reference representation Model 6 S Dynamica Contents...
Страница 7: ...7 Graphical reference representation Model 18 Contents...
Страница 8: ...8 Graphical reference representation Model 24 Contents...
Страница 9: ...9 Graphical reference representation of common components 40 Contents...
Страница 10: ...10...
Страница 63: ...63 Diagrams and wirings...
Страница 75: ...75...