If messages and alarm appear after emptying the tanks, remove the probes and clean
the probes carefully as follows:
- Remove the cover and the faston
(A)
from the
probe bulb.
- Use a wrench
ch 13
to loose the probe holder
(B)
and remove the probe rod.
- Remove the Teflon ogive seal
(C)
; this is generally
deformed to seal the probe holder and assumes the
shape of the threaded hole. To remove the ogive
seal, unscrew the ogive with a screwdriver and
replace it, if required.
- Carefully clean inside the casing
(C)
and the rod
and refit.
Safety valve check
The safety valve should be checked every six months. For this check
chamber
09
should
be cold and
door
16
open.
To ensure proper operations of
safety valve
12
, loosen the cover and unscrew the valve
upper plug. Pull the plug upward to open the valve manually. Close the plug.