28
MESSAGES
Messages contain information and warnings and usually require the user to perform a
simple correction or check.
Message /
Cause
/
•
Solution
Scheduled maintenance
This message indicates the need for a scheduled maintenance on one or more
components.
This message will appear when the steriliser is in STAND BY mode and the remaining cycles
or days of life of one or more components are expiring.
•
The steriliser may however allow to run one or more cycles but the component
subject to scheduled maintenance must be replaced (see chapter MAINTENANCE).
Check water quality
This message appears only on models fitting an external water demineralisation system
(optional) or if the water is automatically loaded from an external tank (optional; standard
on BHD models); in this case, the steriliser monitors conductivity of loaded water.
The steriliser will display this message in STAND BY mode and only when an automatic water
inlet
29
is installed (see USE AND MAINTENANCE MANUAL, chapter CONNECTIONS) and Self-
filling in the Main Settings Menu is ON. This indicates that the water conductivity has
exceeded the optimal level (15 µS/cm) but is still acceptable.
•
The steriliser may allow to run one or more cycles but it is advisable to proceed as
described in section INSUFFICIENT WATER QUALITY because this alarm could stop
operation of the steriliser.
Note: If the water conductivity increases, it means that mineral salts contents has increased
and water is therefore NO LONGER suitable.
Drain waste water
This message indicates that the
w
aste water recovery tank
11
is full and must be emptied.
The steriliser shows this message in STAND BY mode.
•
Empty the tank (see USE AND MAINTENANCE MANUAL, chapter MANUALLY DRAIN
TANKS).
ATTENTION:
The liquid may be very hot and infected and must be disposed
according to local regulations.
•
If the
direct drain
30
is installed, you should check that it is connected correctly
and that the pipe is neither bent nor clogged.
•
If the message is still displayed after emptying the tank, check:
- the level probe which could be dirty or fouled by algae or lime scale caused by
water stagnating in the tank; this could generate false readings (see LEVEL PROBE
TEST);
Messages
Содержание STERILCLAVE 18 B
Страница 2: ...2...
Страница 6: ...6 Graphical reference representation Model 6 S Dynamica Contents...
Страница 7: ...7 Graphical reference representation Model 18 Contents...
Страница 8: ...8 Graphical reference representation Model 24 Contents...
Страница 9: ...9 Graphical reference representation of common components 40 Contents...
Страница 10: ...10...
Страница 63: ...63 Diagrams and wirings...
Страница 75: ...75...