51
Installation of the SPEEDY WATER demineralization system
The SPEEDY WATER demineralization system should be used to connect the SterilClave
6/18/24 sterilisers to the mains water in automatic water inlet configuration B (see USER
AND MAINTENANCE MANUAL, chapter CONNECTIONS).
The SPEEDY WATER demineralization system consists of:
•
conductivity meter
, to be fitted on the steriliser;
•
purification block B
(ionic exchange resins), easy to substitute;
•
support bracket A
for the above mentioned block, to be attached to the wall.
Control of water quality
The conductivity meter allows the steriliser to constantly check that the quality of water
coming from the purification block is suitable for sterilization.
In particular, when the first threshold (15 µS/cm, maximum conductivity recommended
under EN 13060 for feedwater) is exceeded, when in STANBY mode the steriliser displays
the message “Check water quality”, so as to indicate that it is time to change the resins.
When the second threshold (30 µS/cm) is exceeded, the INSUFFICIENT WATER QUALITY
ALARM stops water coming in automatically from the purification block and warns that it is
necessary to change the resins.
Purification block
By treating mains water with a flow of about 650 µS/cm the
purification block B
is able to
produce 250 liters of water below the first threshold (15 µS/cm) and another 50 before
reaching the second (30 µS/cm) threshold, giving a total of 300 liters (
600 cycles).
Technical data
Flow rate
between 0,5 and 0,7 l/min
Dimensions H x W x D
520 x 160 x 90 mm
Operating weight
4.7 kg
Mains water requirements
pressure between 1 and 4.5 bar
Support bracket
The
support bracket A
is equipped with rapid connections for connecting to the mains
water (inlet) and to the steriliser (outlet). These fittings are suitable for 8 mm external
diameter pipes. There are also four 7 mm holes for wall mounting.
Bracket technical data
Dimensions H x W x D
270 x 160 x 180 mm
Operating weight
0.9 kg
Maintenance of the SPEEDY WATER demineralization system
When the
purification block B
needs to be changed:
•
Switch OFF the steriliser, open the mains water tap and put a container to collect any
water spillage under the
support bracket A
.
• Lift the front cover of the
support bracket A
.
• Remove the
metal blocking plate D
.
• Detach the
purification block B
from the support and install the new one as set out in the
instructions below.
Speedy Water
Содержание STERILCLAVE 18 B
Страница 2: ...2...
Страница 6: ...6 Graphical reference representation Model 6 S Dynamica Contents...
Страница 7: ...7 Graphical reference representation Model 18 Contents...
Страница 8: ...8 Graphical reference representation Model 24 Contents...
Страница 9: ...9 Graphical reference representation of common components 40 Contents...
Страница 10: ...10...
Страница 63: ...63 Diagrams and wirings...
Страница 75: ...75...