47
Proceed as follows for vacuum pump 41 B:
•
Remove the pump and carefully check the position and orientation of
components.
•
Check that membrane
C
, valves
D
and
O-
Rings
are not damaged. clean them or
replace them if required.
•
Membrane
C
is fitted onto the pump piston
by turning it clockwise. To prevent
movements, use thread lock liquid or paste.
•
When refitting, be careful to arrange the
parts correctly. It may not be possible to refit
the parts if they are not correctly oriented.
Refit outlet
F
after fastening membrane
C
.
Slide the lips of membrane
C
over the inner
ring of the block using a screwdriver.
Note:
If the pump is to be replaced, remove the
rubber plugs
E
.
Note:
Check correct fastening of the delivery and
suction fittings.
Vacuum pump kit 41 B:
Vacuum pump 41 B:
Additional Checks And Settings