19
Using the cable included in the kit, connect connector CN4 of the calibration tool to
connector CN11 of the
power board
34
.
Set the calibration tool switch to 0 Vdc.
Wait for at least 10 seconds and press ENTER.
Now set the calibration tool switch to 5 Vdc.
Wait for at least 10 seconds and press ENTER.
Press to go back to the Technician Menu without changing the calibration.
Technician Menu
Содержание STERILCLAVE 18 B
Страница 2: ...2...
Страница 6: ...6 Graphical reference representation Model 6 S Dynamica Contents...
Страница 7: ...7 Graphical reference representation Model 18 Contents...
Страница 8: ...8 Graphical reference representation Model 24 Contents...
Страница 9: ...9 Graphical reference representation of common components 40 Contents...
Страница 10: ...10...
Страница 63: ...63 Diagrams and wirings...
Страница 75: ...75...