Disconnect the cable from connector CN1 of the calibration tool and connect connector
CN2 of the calibration tool (1573 Ohm sample coil, corresponding to 150°C) to calibrate
the sensor PT1 or the connector CN3 (1832 Ohm sample coil, corresponding to 220°C) to
calibrate sensors PT2 and PT3.
Wait for at least 10 seconds and press ENTER.
Press to go back to the Technician Menu without changing the calibration.
Technician Menu