OWNERS MANUAL | Cobolt Skyra™ |
D0423-E JANUARY 2018
5 | 45
Safety
General
The laser classification of Cobolt Skyra™ is dependent on the specific laser lines present a multi-line laser. The
wavelengths and maximum accessible emission can be found on the warning label on the laser head. See section 2.1
for detailed information on warning and identification labels.
The sum of the maximum emitted power of all laser lines in the multi-line laser determines the laser classification.
The residual emission at wavelengths outside the specified range do not exceed Laser Class 1. A multi-line laser
capable of emitting less than 500 mW of laser radiation within the visible spectrum (405 nm – 660 nm) is Class IIIB
(CDRH), Class 3B (IEC).
A multi-line laser capable of emitting more than 500 mW of laser radiation within the visible spectrum (405 nm – 660
nm) is Class IV (CDRH), Class 4 (IEC).
Eye and skin exposure to direct or reflected laser light is hazardous and may be extremely harmful. Always wear eye
protection appropriate to the beam wavelength(s) and intensity and never look directly into a laser beam. Laser
radiation may ignite flammable materials and combustible gasses in the beam path and, in event of ignition, fumes
may be generated. All equipment used in close proximity to the laser beam should be suitably fire resistant and the
facility should be properly ventilated. It is advised to perform a risk assessment for the facility and equipment prior to
using the laser. In the case of integration into a larger system, laser safety compliance must be evaluated in the end
product. The device must be handled by skilled personnel experienced with lasers, in a laboratory environment and
with access to adequate laser safety equipment. The laser head clearly displays a yellow warning label that shows the
location of the laser beam aperture. This label must be visible unless the laser beam is totally enclosed. If the laser
does not function, do not attempt to open any of the units, or the warranty will be voided. Call or e-mail your local
Cobolt representative for consultancy and to request an RMA number (see back cover for contact information).
CAUTION Use of controls or adjustments or performance of any procedures other than those
specified herein may result in exposure to hazardous radiation.