Compliance (CDRH models only)
The CDRH model lasers (-1/300) are designed and manufactured to comply with the EC Low Voltage Directive and the
EC EMC Directive in the CDRH-compliant configuration of laser head, key control box, key and Cobolt-supplied power
supply. All equipment must be mounted on a common ground plane, such as an optical table. If any part of the
delivered equipment is replaced with a part not supplied by Cobolt or if the equipment is not properly grounded
system may not conform to CE / CDRH compliance standards listed in section 15. Disabling any of the safety features
nullifies the CE marking and violates the laser safety standard.
The following harmonized and limits standards have been applied:
Electrical Safety:
EN 61010-1, IEC-61010-1, UL 61010-1 (Limited Energy System)
Laser Safety/Class
IEC-60825-1 , CDRH 21 CFR 1040.10 and 1040.11
EMC
4-2
IEC 61326-1
EN 55011
Electromagnetic Emission , Class B
FCC Part 15, subpart B, class B
Electromagnetic Immunity – Table 2 Requirements
EN 61000-4-2
Electrostatic Discharge
±4 kV contact discharge and
±2 kV, ±4 kV, ±8 kV air discharge
EN 61000-4-3
Radiated electromagnetic fields
80 – 1000 MHz 10 V/m with 80 % AM @ 1 kHz
1.4 – 2.7 GHz 10 V/m with 80 % AM @ 1 kHz
EN 61000-4-4
Fast transient / Burst
AC Power input port ±2,0 kV
EN 61000-4-5
Surge
AC Power input port ±0,5 kV, ±1,0 kV, ±2,0 kV Com. Mode
AC Power input port, ±0,5 kV, ±1,0 kV Diff. Mode
EN 61000-4-6
Conducted Immunity
10 V with 80 % AM @ 1 kHz
EN 61000-4-11 Dips and Interruptions
50 Hz and 60 Hz. Test voltages: 100 V and 230 V
RoHS
EU Directive 2011/65/EU
Contact your sales representative for a copy of the full Declaration of Conformity.