All safety recommendations in section 2 are also valid for the Cobolt Skyra™ fiber coupled laser heads. Additionally,
heat generated from absorption of laser radiation by particles on the fiber end may increase the probability of ignition
hazards in certain environments. Always clean the fiber end before turning on the laser. In systems where the beam
is exposed, fiber end must be mounted < 2 m from the emission warning LED. It is advised to perform a risk
assessment for the facility and equipment prior to using the laser. In the case of integration into a larger system, laser
safety compliance must be evaluated in the end product.
Accessible emission information for Cobolt Skyra™ laser lines
The table below describes the irradiance in W/cm
2
and appropriate level of eye protection in terms of optical density
(OD) for each product line. Using a multi-line laser requires a combination of eye protection that is specifically tailored
to the device configuration. Be sure to consider all available wavelengths and power levels when selecting eye
protection.
Laser line
Nominal Output Power
(mW)
Max Power
(mW)
Max Irradiance
(W/cm
2
)*
Eye protection Requirement
(OD)**
Cobolt MLD 405 nm
50
120
50
4
100
120
50
4
Cobolt MLD 445 nm
50
120
42
4
Cobolt MLD 473 nm
50
120
42
4
Cobolt MLD 488 nm
50
80
28
3
Cobolt MLD 515 nm
50
100
35
3
Cobolt DPL 532 nm
50
100
150
300
53
106
3
3
Cobolt DPL 553 nm
50
100
32
3
Cobolt DPL 561 nm
50
150
48
3
Cobolt MLD 633 nm
50
100
35
3
Cobolt MLD 638 nm
50
120
71
3
Cobolt MLD 647 nm
50
150
53
3
Cobolt MLD 660 nm
50
130
46
3
* Irradiance (W/cm2) = Max Power (W)
Beam Area at bottom tolerance (cm2)
** Eye protection (OD) = Log
10
( Max Power (W)
60825-1 Emission Limit : Class 1 (W) ) , rounded up to the next integer.