OWNERS MANUAL | Cobolt Skyra™ |
D0423-E JANUARY 2018
28 | 45
Pin assignment
All equipment connected to the system should be limited energy as described by IEC 61010:1.
Laser head
The pin configuration for the 10 pin Molex connector on the laser head is described in the table below.
Pin
Signal
1
Remote interlock
2
0
V – Ground
3
Direct On/Off (+5 V Input) – OEM Only
4
Key Switch
5
LED 1A (Laser On)
6
LED 2 (Error)
7
LED 1B (Laser On - Redundant)
8
Not used (test)
9
Not used (Ground)
10
Not used (Ground)
Key box
The pin configuration for the 15-pin D-SUB on the key control box are described in the table below.
Pin
Function
1
LED1 (A and B) - Laser on
2
LED2 - Error
3
Not used
4
0
V (ref pin 1,2,5,11)
5
Key Switch
6
Not used
7
Not used
8
Not used
9
Not used
10
0
V GND (ref pin 5)
11
Remote interlock connector
12
Not used
13
Not used
14
Not used
15
Not used
1 3 5 7 9
2 4 6 8 10
10 pin Molex socket on laser head