Low power
1.
Check that the laser is in constant power mode (using the GUI or the “cp” command).
2.
Check the power reading using the GUI or the “pa?” command.
3.
If this does not agree with the real output power, re-calibrate by measuring the power and entering it in the
“Power Cal” box in the software.
4.
Send the command “f?” If fault code 4 is returned, there may be a problem with the constant power system.
5.
Contact Cobolt technical support.