OWNERS MANUAL | Cobolt Skyra™ |
D0423-E JANUARY 2018
13 | 45
Cobolt Monitor™ software expanded to for more detailed monitoring.
TEC Settings
Shows the running status and the fault status for the laser’s internal thermoelectric coolers (TEC).
Laser Operation Mode and Settings
Displays the set laser power. The user can switch between constant power mode, constant current
mode and modulation mode. Likewise, there are boxes to set the constant power level and constant
current level. In constant power mode the current will be set by Cobolt Monitor to reach the power
level set in the this field. When in modulation mode it is recommended to use an external power
meter, the internal measurement will not be reliable. See section 8 and 9 for more details on
continuous wave and modulation modes. Use the tabs to go from laser to laser.
Autostart Program
Displays whether the laser is in CDRH or OEM mode and displays the current laser operational
status. There are also buttons to “abort” the autostart sequence or to “restart” the laser after a fault.
Waiting for Temp
Laser emission is not enabled until all temperatures have reached their set point and
the TECs are stabilized.
Waiting for Key
Toggle the key to proceed, if the key is in the ON position, turn of OFF and ON again.
Warm up
A current is provided to warm the laser before setting full power
Completed
The device is emitting or armed for emission.
Fault
The device has a fault, the fault status must be cleared before the laser can be
restarted.
Aborted
The autostart sequence has been aborted, but the TECs are still running.