3.10 REQUIRED QUARTERLY PREVENTIVE MAINTENANCE CHECKLIST
HEMOTHERM Model 400CE Serial No. _________________________
Check when
Hospital I.D. No. _________________________
completed
1.
External cabinet and controls in good condition (i.e.
2.
All warning labels properly affixed.
3.
Quick disconnect couplings are tight, straight, and not leaking.
4.
Power cord (i.e. no cuts or exposed wire) and plug (i.e. no bent or missing pins)
are in good condition.
5.
All indicator lights operative (heat & cool modes, heaters, pump, power).
(Press TEST INDICATORS BUTTON on keypad, Section 1.3)
6.
Drain and clean reservoirs (See Section 3.4).
7.
Clean water system (See Section 2.4).
8.
Clean water filter (See Section 3.6).
9.
Refill reservoirs with distilled water (See Section 3.4).
10.
Clean condenser and grille (See Section 3.7).
11.
Leakage current check (all readings should be less than 300 micro amps for 115V
units and 500 micro amps for 230V units).
(See Op/Tech Manual)
OFF normal polarity
OFF reverse polarity
ON normal polarity (heat)
ON reverse polarity (heat)
ON normal polarity (cool)
ON reverse polarity (cool)
12.
Perform low limit safeties check.
(See Op/Tech Manual)
13.
Perform high limit safeties check.
(See Op/Tech Manual)
14.
Check condition of blankets, hoses, couplings (check for leaks).
(See Section 3.8)
_________________________________
__________________
Signature of Inspector
Date of Inspection
Содержание HEMOTHERM 400CE
Страница 1: ......
Страница 47: ...HƯỚNG DẪN SỬ DỤNG MÁY TRAO ĐỔI NHIỆT HEMOTHERM MODEL 400CE Trang 47 46 ...
Страница 64: ...OPERATION MANUAL HEMOTHERM MODEL 400CE Page 17 of 46 FIGURE 1 2 KEYPAD 115V UNIT FIGURE 1 3 KEYPAD 115V REMOTE ...
Страница 65: ...OPERATION MANUAL HEMOTHERM MODEL 400CE Page 18 of 46 FIGURE 1 4 KEYPAD 230V UNIT FIGURE 1 5 KEYPAD 230V REMOTE ...
Страница 94: ...OPERATION MANUAL HEMOTHERM MODEL 400CE ...