OPERATION MANUAL
HEMOTHERM MODEL 400CE
Page
43
of
46
1.
Press the power switch to the OFF/”O” position.
2.
Disconnect the unit from its power source.
3.
Using a Phillips head screw driver, remove the five screws of the rear
enclosure panel. Guide the panel outward towards the floor.
4.
Locate the water filter assembly.
5.
Firmly grasp the notched rim of the cap of the assembly and turn it
counterclockwise. The cap and hose may contain a small amount of water.
The wire mesh may be lodged in the top of the fitting or it may be resting in
the plastic cap.
6.
Remove the wire mesh.
7.
Clean the wire mesh and the plastic cap. Be careful not to lose the black O-
ring in the rim of the plastic cap.
8.
Replace the wire mesh in the plastic cap and position the wire mesh, O-ring
and plastic cap under the fitting.
9.
Screw the cap clockwise on to the fitting until it is secure.
10. Replace the rear enclosure. Replace the five screws.
3.7
MAINTENANCE OF THE CONDENSER GRILLE
Cool air is taken in through the grille at the front of the HEMOTHERM unit. Warm air is
evenly expelled through the bottom of the unit. Both the air intake and outlet must be
kept clear when the machine is in operation.
If the condenser grille becomes covered with dust and lint, the cooling capacity of the unit
will be reduced. AT LEAST EVERY THREE MONTHS THE CONDENSER SHOULD BE
CLEANED. To do this, first loosen the four screws securing the grille over the condenser
intake. Then, remove accumulated dust or lint with a brush or vacuum. Finally, replace
grille cover and tighten all four screws.
3.8
MAINTENANCE OF THE HYPER-HYPOTHERMIA BLANKETS
This section describes the general maintenance and for the reusable and disposable
blankets. General maintenance tasks include cleaning, draining, and storing the blankets.
For further maintenance information on Cincinnati Sub-
Zero blankets refer to the blanket’s
corresponding Instructions for Use.
Содержание HEMOTHERM 400CE
Страница 1: ......
Страница 47: ...HƯỚNG DẪN SỬ DỤNG MÁY TRAO ĐỔI NHIỆT HEMOTHERM MODEL 400CE Trang 47 46 ...
Страница 64: ...OPERATION MANUAL HEMOTHERM MODEL 400CE Page 17 of 46 FIGURE 1 2 KEYPAD 115V UNIT FIGURE 1 3 KEYPAD 115V REMOTE ...
Страница 65: ...OPERATION MANUAL HEMOTHERM MODEL 400CE Page 18 of 46 FIGURE 1 4 KEYPAD 230V UNIT FIGURE 1 5 KEYPAD 230V REMOTE ...
Страница 94: ...OPERATION MANUAL HEMOTHERM MODEL 400CE ...